Climatic Chambers
Climatic chambers are available in the ESE group to allow the testing of electronics modules at different temperatures. They are appropriate for temperature and humidity cycling tests for quality verification of board assemblies. Note that ESE is not equipped to perform burn-in testing of large production batches.
The climatic chambers in ESE are located in 13-S-011. The technical details are summarised below. The full list of available chambers, their technical details, the planning and reservation form are available here:
https://epool.cern.ch/climatic-chamber
Contact:


Climats EXCAL 1423-HA (2013):
Temperature range: -80 C, +180 C (±1°C after stabilization). Interanal probe Limit: 155°C
Temperature rate (heating) de -70°C à +180°C: : 17°C/min
Temperature rate (cooling) de +180°C à -70°C: 8°C/min
Relative humidity range (Dewpoint limit +6°C) 10% at 98% (+10°C to +90°C)
Volume : 138l, 55cm * 50cm * 50 cm (L * l * H)
Signals to the outside: Max window for cables: hole of 8cm diameter
Please note that humidity may imply high water consumption, so it will be necessary to foresee water purchase: SCEM: 58.23.10.200.8
CTS Climatic Chamber Type T-40/500 (2004):
Temperature range: -40 C, +180 C
Temperature rate (heating): : 1.3 K/min
Temperature rate (cooling): 1.2 K/min
Humidity: Simple air dryer
Volume: 500 liters , 65cm x 65cm x 120cm (L x l X h)
Control: Manual or Programmable via RS232
Signals to the outside: Max window for cables: hole of 8cm diameter